ASTM D3895 ISO 10837 Oxidative Induction Time Tester DW1470 is used to measure the oxidative induction time (OIT) of plastic or rubber materials.
DW1470
FYI
white and black
1PCS
China
By TT or LC
15-20 days
Availability: | |
---|---|
Application
ASTM D3895 ISO 10837 Oxidative Induction Time Tester DW1470 is used to measure the oxidative induction time (OIT) of plastic or rubber materials. The time is used as an indication of the oxidative thermal stability of plastic or rubber materials. When samples are placed into the rack of the instrument and protected in Nitrogen, heat them at the rate of 20℃/min until the temperature reaches 200℃±0.1℃; Hold the temperature for 7 minutes and then quickly switch Nitrogen to Oxygen. Samples will react with Oxygen to produce heat energy. Curve of change in temperature will be recorded and the time when there is an obvious change on the curve is regarded as the oxidative induction time.
Features OF ASTM D3895 ISO 10837 Oxidative Induction Time Tester
1. Integrative design with less signal loss and less interference;
2. Small-scale heating furnace with less thermal inertia to linearly heat samples from the beginning of room temperature and more precise constant-temperature control;
3. Two thermometers: one to real-timely display temperature in the heating furnace; the other to display temperature on samples during test;
4. Perfect air control system consisting of two air passages to control air flow in quality and automatic system to switch Nitrogen to Oxygen during test;
5. Automatically measure oxidative induction time and protract a curve;
6. Automatically draw the tangent, seek an intersection and get the time length between the intersection and the point where Nitrogen is switched to Oxygen;
7. Large LCD screen to real-timely display state of the instrument and test data;
8. Automatically create test report.
Key specifications
DSC range | ±10, ±20, ±40, ±100, ±160, ±200mW, Automatically switch |
Working temperature | RT ~ 500℃ |
Temperature accuracy | ±0.1℃ |
Heating speed | 1~30℃/min |
Specification of crucible | Aluminum, Φ6.7mm×2.5mm |
Gas flow rate | ≤200ml/min |
Gas pressure | 0.2Mpa |
Microcomputer data processing system | Data acquisition and storage, LCD display, test report, calibration and correction of temperature, process of oxidation induction time |
Power supply: | AC 220V, 50Hz (Ensure reliable grounding. The power supply must be maintained stable with no frequent fluctuation, in order to avoid any influence on the test result.) |
Ambient conditions: | Temperature: 15~25℃, Humidity: 55~75%RH No obvious vibration or interference. |
Application
ASTM D3895 ISO 10837 Oxidative Induction Time Tester DW1470 is used to measure the oxidative induction time (OIT) of plastic or rubber materials. The time is used as an indication of the oxidative thermal stability of plastic or rubber materials. When samples are placed into the rack of the instrument and protected in Nitrogen, heat them at the rate of 20℃/min until the temperature reaches 200℃±0.1℃; Hold the temperature for 7 minutes and then quickly switch Nitrogen to Oxygen. Samples will react with Oxygen to produce heat energy. Curve of change in temperature will be recorded and the time when there is an obvious change on the curve is regarded as the oxidative induction time.
Features OF ASTM D3895 ISO 10837 Oxidative Induction Time Tester
1. Integrative design with less signal loss and less interference;
2. Small-scale heating furnace with less thermal inertia to linearly heat samples from the beginning of room temperature and more precise constant-temperature control;
3. Two thermometers: one to real-timely display temperature in the heating furnace; the other to display temperature on samples during test;
4. Perfect air control system consisting of two air passages to control air flow in quality and automatic system to switch Nitrogen to Oxygen during test;
5. Automatically measure oxidative induction time and protract a curve;
6. Automatically draw the tangent, seek an intersection and get the time length between the intersection and the point where Nitrogen is switched to Oxygen;
7. Large LCD screen to real-timely display state of the instrument and test data;
8. Automatically create test report.
Key specifications
DSC range | ±10, ±20, ±40, ±100, ±160, ±200mW, Automatically switch |
Working temperature | RT ~ 500℃ |
Temperature accuracy | ±0.1℃ |
Heating speed | 1~30℃/min |
Specification of crucible | Aluminum, Φ6.7mm×2.5mm |
Gas flow rate | ≤200ml/min |
Gas pressure | 0.2Mpa |
Microcomputer data processing system | Data acquisition and storage, LCD display, test report, calibration and correction of temperature, process of oxidation induction time |
Power supply: | AC 220V, 50Hz (Ensure reliable grounding. The power supply must be maintained stable with no frequent fluctuation, in order to avoid any influence on the test result.) |
Ambient conditions: | Temperature: 15~25℃, Humidity: 55~75%RH No obvious vibration or interference. |