Protective Clothing Blood Penetration Resistance Tester DW0521 is used to determine the resistance of medical protective clothing to the penetration of synthetic blood under specified test conditions.
DW0521
FYI
Black
1PCS
China
Availability: | |
---|---|
APPLICATION:
Protective Clothing Blood Penetration Resistance Tester DW0521 is used to determine the resistance of
medical protective clothing to the penetration of synthetic blood under specified test conditions.
RELATED STANDARDS:
STANDARDS | ISO 16603 | ASTM F1670 | ASTM F1671 |
Note: this tester can conform to but not limit to the standards above, for more standards conformance, please contact us.
FEATURES:
1.Color touch screen display and operation;
2.The test tank is processed with high quality PTFE material, and the top is equipped with a transparent protective cover;
3.The test pressure can be adjusted. It can keep pressurizing under different pressures;
4.Clamping force is accurate and reliable;
5.The timer displays the test time in real time.
KEY SPECIFICATION:
Specimen size | 75*75mm |
Test pressure | 1.75kPa, 3.5kPa, 7kPa, 14kPa, 20kPa Or can adjust between 0-40 kPa |
Timer | 0.1-999.9s, resolution 0.1s |
Retaining screen | 304 stainless steel, aperture ratio≥50%, bending ≤5mm at 14kPa |
Power Supply | AC220V, 50Hz |
Dimensions | 500×420×340 mm (L×W×H) |
APPLICATION:
Protective Clothing Blood Penetration Resistance Tester DW0521 is used to determine the resistance of
medical protective clothing to the penetration of synthetic blood under specified test conditions.
RELATED STANDARDS:
STANDARDS | ISO 16603 | ASTM F1670 | ASTM F1671 |
Note: this tester can conform to but not limit to the standards above, for more standards conformance, please contact us.
FEATURES:
1.Color touch screen display and operation;
2.The test tank is processed with high quality PTFE material, and the top is equipped with a transparent protective cover;
3.The test pressure can be adjusted. It can keep pressurizing under different pressures;
4.Clamping force is accurate and reliable;
5.The timer displays the test time in real time.
KEY SPECIFICATION:
Specimen size | 75*75mm |
Test pressure | 1.75kPa, 3.5kPa, 7kPa, 14kPa, 20kPa Or can adjust between 0-40 kPa |
Timer | 0.1-999.9s, resolution 0.1s |
Retaining screen | 304 stainless steel, aperture ratio≥50%, bending ≤5mm at 14kPa |
Power Supply | AC220V, 50Hz |
Dimensions | 500×420×340 mm (L×W×H) |